CT Image Quality Phantom
The CT image quality phantom is designed and built with eight main cylinders to measure CT number accuracy, slice thickness accuracy, geometric accuracy, spatial resolution, point spread function (PSF), edge spread function (ESF), modulation transfer function (MTF), contrast resolution and uniformity.
In this phantom, spatial resolution can be measured in terms of the number of cycles per unit length, the number of pairs of lines per unit length, and the PSF, ESF, and MTF. In addition, in the designed phantom, the spatial resolution can be measured up to 21 pairs of lines per centimeter. Also in this phantom, the contrast resolution consisted of six sets of cylinders with different contrasts and sizes.
The CT image quality phantom is capable of evaluating a wide range of parameters affecting CT image quality. In addition to performing quality control tests, this phantom can be used to optimize noise and dose reduction techniques


Product Features
1- CT Number Accuracy Measurement Module
2- Section Thickness Measurement Accuracy Module
3- Spatial Resolution Measurement Module (Cycles per Unit Length)
Point spread function measurement module, edge spread function, modulation transfer function
Spatial resolution measurement module (up to 21 line pairs/mm)
Contrast resolution measurement module
Geometric dimension accuracy measurement module
Image uniformity measurement module
Storage compartment
Handle
Horizontal level along the y-axis
Horizontal level along the x-axis
Feet
Storage compartment door
:Specifications of different CT scan quality control phantom modules
| CT number accuracy measurement module:
|
||
| Dimensions and structure | A cylindrical piece with a diameter of 198 mm, with eight cylinders with diameters of 25 mm, made of tissue-equivalent material.
|
![]() |
| Checkable parameters | Accuracy and repeatability of CT or Hounsfield numbers of tissue equivalent materials
|
|
| Measurement range |
numbers include: 5/4 + ، 106+ ، 72- ، 26- ، 89+ ، 123+ ، 951+ و 997-
|
|
| material | Acrylic and equivalent textured materials | |
| Section thickness accuracy measurement module:
|
||
| Dimensions and structure | A cylindrical piece with a diameter of 198 mm consisting of four stainless steel wires at an angle of 23 degrees.
|
![]() |
| Checkable parameter | The correct size of the thickness of the section
|
|
| Measurement range | From one tenth to 15 mm
|
|
| material | Acrylic body and stainless steel wire included | |
| Spatial resolution measurement module (in terms of number of cycles per unit length):
|
||
| Dimensions and structure | A cylindrical piece with a diameter of 198 mm containing three sets of eight rows of holes of different diameters.
|
![]() |
| Checkable parameter | Spatial resolution in terms of number of cycles per unit length
|
|
| Measurement range |
diameter of the holes include 75/1 ، 5/1 ، 25/1 ، 1 ، 75/0 ، 61/0 ، 50/0 و 40/0 eqaul to mm 86/2، 33/3، 4، 5، 67/6، 20/8، 10، 5/12 Line pairs per centimeter
|
|
| material | acrilyc | |
| Point Spread Function (PSF), Edge Spread Function (LSF), Modulation Transfer Function (MTF) Measurement Module:
|
||
| Dimensions and structure | A cylindrical piece with a diameter of 198 mm has a Teflon half-cylinder and an empty cylinder and a stainless steel wire with a diameter of 0.25 mm in the middle.
|
![]() |
| Checkable parameter | PSF، LSF, MTF
|
|
| Measurement range |
From zero to one hundred percent for MTF |
|
| material | acrylic | |
| Spatial resolution measurement module (in terms of number of line pairs per unit length):
|
||
| Dimensions and structure |
A cylindrical piece with a diameter of 198 mm containing 21 sets of grooves of different widths
|
![]() |
| Checkable parameter |
Spatial resolution in terms of number of line pairs per unit length
|
|
| Measurement range | From 1 to 21 number of line pairs per centimeter | |
| material | acrylic | |
| Contrast resolution measurement module:
|
||
| Dimensions and structure |
A cylindrical piece of acrylic with a diameter of 198 mm, consisting of six sets of cylinders with different contrasts and sizes, including three sets of cylinders around the perimeter with diameters of 15 ،9 ، 8، 7، 6، 5، 4، 3، 2، 1 و 5/. Three sets in the center with diameters 9، 7، 5 , 3
|
![]() |
| Checkable parameter |
|
|
| Measurement range | Contrast resolution
|
|
| material | Acrylic and fillers | |
| Geometric accuracy measurement module:
|
||
| Dimensions and structure |
Cylindrical piece with diameter 198 8holes with 4 circles in 6 diameter In hypothetical positions of the clock 12، 3، 6 , 9 4circles in 4 mm diameter in hypothetical position of the clock 5/1,5/4, 5/7 , 5/10
|
![]() |
| Checkable parameters | Geometric accuracy
|
|
| measurment range |
100mm between consecutive large or small holes
|
|
| material | acrylic | |
| Image uniformity measurement module:
|
||
| dimansion and structure |
Cylindrical container with a diameter of 198 mm 34 mm wide with two valves for filling or emptying distilled water
|
![]() |
| Checkable parameters | Image uniformity and artifacts
|
|
| Measurement range | CT numbers
|
|
| material | Acrylic cylindrical container containing a uniform substance (distilled water or air)
|
|
Compatible with:
– IEC 61223-3-5
– IEC 61223-2-6
– AAPM guideline